D.E. Nowak
Materials Science and Engineering, Cornell University
 

Abstract:
A new multipurpose diffractometer is being designed in collaboration with Advanced Design Consulting (ADC) for use in the G2 hutch. A six circle, kappa style, design was chosen to maximize versatility and eliminate obstacles, such as an Eulerian cradle, to incident or diffracted x-ray beams. The design will allow the use of a relatively large sample stage that weighs 7 kg (15.4 lb). The diffractometer consists of two independent goniometers, a four-circle kappa goniometer and a two-circle two-theta goniometer. The kappa goniometer is used to position the sample in any orientation with respect to the incident beam without blocking the diffracted beam. The two-theta goniometer consists of two perpendicular circles that can position a detector at any point on the Ewald sphere.

2008 Run